​Analysis of common problems in the verification of electronic floor scales

Keyword:electronic floor scales   Time:2019-11-12 9:20:32

Analysis of common problems in the verification of electronic floor scales

 1.1  problems of scale itself

 Weighing platform is an important component of electronic floor scale, which mainly plays an important role in carrying heavy objects. When the electronic floor scale is used for verification, it is necessary to check the bottom surface of the symmetrical body, the weighing body itself and the clearance of the protective frame at both ends of the weighing body to see whether there are obvious problems. There are several common problems of weighing body: whether there is debris stuck weighing body, whether there is not enough compact connection between weighing body, whether there is displacement of weighing body sensor, etc.

 1.2  sensor problem.

 The sensor is an important component between the base and the weighing platform. The main function of the sensor is to convert the weight of the cargo into the corresponding electrical signal. The sensor damage may be caused by the said goods being seriously overweight, or it may be caused by the said goods being deficient in quality. Once the sensor is damaged, there will be poor insulation performance, zero point drift or signal line loose, such as the fault.

 1.3  the testing technology is not up to standard, and the obtained test results have a large error.

 The accuracy of partial load test results determines the accuracy of the final test results. However, in actual measurement, due to the difference of weight specifications and models or the existence of standard difference, it is easy to cause the deviation of the obtained test results from the actual situation. This phenomenon of inaccurate partial load test has a direct impact on the verification of the equipment to meet the actual requirements.

 1.4  instrument problems

 The function of electronic scale display instrument is to measure the current and voltage change of the sensor, so as to measure the weight accurately. If you want to find out exactly where the failure of the electronic weighmeter occurs, you can usually use the simulator or an alternative method to diagnose the failure of the meter. When the method of substitution is used in the measurement, a substitute which is close to one half of the maximum weight can be used to reload the load carrier three times to check its repeatability.

 1.5  limited testing conditions and imperfect metering infrastructure

 In China, there is still insufficient capital investment in metrological infrastructure construction, which leads to the limited verification conditions of metrological verification units in some areas, and the lack of standard instruments generally exists, which is not conducive to meeting the needs of practical work. Combined with the current actual verification departments at the county (city, district) level, the weight of the verification equipment is usually less than or equal to 50t, but the verification operation of >50t electronic floor scales frequently occurs in practice, which is seriously inconsistent with the current practical needs. At the same time, safe transportation of standard weights is also a very important and difficult problem. Safe transportation standard weight not only needs to bear high transportation costs but also virtually increases the cost input. It is necessary to ensure the rigor of standard weight transportation and try to avoid safety accidents caused by bumpy shift during transportation.

 When testing a scale, the closer the test weight is placed to the sensor, the better the support position.

 In addition, as there are many types of electronic floor scales on the market, different manufacturers will usually adopt different calibration procedures if there is a problem of biased load. For analog sensor electronic floor scale, deviation of partial load can usually be solved by using potentiometer. For digital sensors, the deviation can be directly corrected by the meter head. Shantou quality measurement supervision and inspection, for example, is used by static electronic floor scale, measured on the device performance test and cycle test, usually in accordance with the following criteria: the in-car total quality inspection criterion car there are 16 tons the following two specifications, one of the specifications and weight of 1 t, another  specification weight is 500 kg. In the actual verification operation, if the total weight of the floor scale is within the weight range of the standard weight, the basic operation steps can be used for verification.